Development and production:

Description

Digital pulse processing (DPP) systems are known to have better performance than analog ones. DPP can synthesize almost any pulse response shape without the associated signal degradation which happens in a complex analog path.

This board contains high-speed transient recorders with auto-trigger functionality, used to digitize and store the detailed shape of pulses. Algorithms for the detection of pulse pile-up, Pulse Height Analysis, and, optionally, Pulse Shape Discrimination are implemented on the on-board FPGA, and can be applied to the digitized pulses for data reduction or real-time spectra monitoring.

Characteristics

  • ATCA board with eight acquisition channels per module;
  • Data transfer rate of up to 800 Mbyte/s over x4 PCI Express to the host processor.
  • A maximum sampling rate / resolution of 250 MSPS @ 13-bit, 400 MSPS @ 14-bit or 500 MSPS @ 12-bit depending on the ADC choice;
  • The channels can be paired for a higher sampling rate, respectively: 500, 800 or 1000 MSPS; the ENOB figure is then as low as ~10-bit;
  • Maximum pulse rate in excess of 5 Mpulse/s are attainable(e.g. 6.25 MSPS for 400 MSPS @64 sample/pulse) since there is no dead-time during data processing;
  • ENOB higher than11-bit (measured for energies from 4 MeV to 20 MeV);
  • 2 Gbytes of local memory (DDR2) arranged as two memory modules, each one storing the acquired data of four ADC channels (256 Mbytes/channel);
  • Digitally programmable pulse level detection with configurable amount of pre-trigger and pos-trigger samples;
  • Optional programmable input voltage offset to configure for maximum signal amplitude on the analogue input (-1V to +1V @ 50 ohms);
  • Internall or external start trigger (in the front-panel);
  • Externally/internally clocked/triggered programmable time mark (1 ms nominal) to trigger the storage of the pulse counting value. The acquired pulse vector between two marks can be used to calculate lap spectra;
  • The local timestamp has a resolution as low as 1.25 ns and spans, at least, one day;

 

sratca.png

User manual

User manual base document.

Annex A - Setting an acquisition on the Linux command line.

Annex B - Board registers.

Was this page helpful?
Tag page (Edit tags)
  • No tags

Files 8

FileSizeDateAttached by 
 AnnexA.pdf
No description
715.64 kB19:42, 23 Jan 2012jsousaActions
 AnnexB.pdf
No description
584.07 kB19:42, 23 Jan 2012jsousaActions
 FUSION_4274.pdf
ATCA data acquisition system for gamma-ray spectrometry - FUSION ENGINEERING AND DESIGN, VL 83, IS 2-3, pg:341-345, 2008
473.82 kB18:11, 10 Feb 2011AdminActions
 Fusion_Andre.pdf
ATCA data acquisition system for gamma-ray spectrometry - FUSION ENGINEERING AND DESIGN, VL 83, IS 2-3, pg:341-345, 2008
339.8 kB18:11, 10 Feb 2011AdminActions
 sratca.png
No description
1118.16 kB13:51, 5 Apr 2011jsousaActions
 Tardochi_paper_Rv_scientific_Instrum.pdf
ATCA data acquisition system for gamma-ray spectrometry - FUSION ENGINEERING AND DESIGN, VL 83, IS 2-3, pg:341-345, 2008
226.46 kB18:11, 10 Feb 2011AdminActions
 TCHTPD_rita.pdf
ATCA data acquisition system for gamma-ray spectrometry - FUSION ENGINEERING AND DESIGN, VL 83, IS 2-3, pg:341-345, 2008
163.65 kB18:11, 10 Feb 2011AdminActions
 User_Manual_base.pdf
No description
3.05 kB20:53, 23 Jan 2012jsousaActions
You must login to post a comment.
Powered by MindTouch Core